Installation and operation manual – Brooks Instrument MT3809E User Manual

Page 42

Advertising
background image

3-2

Section 3 Operation

Models MT 3809 and 3819

Installation and Operation Manual

X-VA-MT3809-3819-eng

Part Number: 541B049AHG

September, 2008

3-3 Operation of the Models MT 3809 and MT 3819 Flowmeters with a Smart Meter Manager Transmitter

with or without Optional Alarms and Pulse Output for totalization

a. Start-up the meter as described in Section 3-2.
b. Programming performed prior to shipment (parameters)

The SMM transmitter with or without alarms and pulse output is
preprogrammed prior to shipment for the following parameters based
on the order information, meter configuration and application.
If parameters are not specified in the customer purchase order, the
defaults inherent to the electronics will be as shown in (parenthesis).

Transmitter Only (uP)

Transmitter (uP) Plus...

Pre-Programmed Parameters

Alarms and Pulse Output
Preprogrammed Parameters

Model Number

Alarm LoLimit (0% FS flow)

Serial Number

Alarm HiLimit (100% FS flow)

Tag Name (blank)

Pulse output units (1 ppg)

Flow Rate units of measure (gpm)

Resettable totalizer units of measure (gal.)

Low Flow cutoff (10% FS flow)

Inventory totalizer units of measurement

Calibration factor (% scales only)

(gal.)

Analog Output LoRange

(4 mA @ 0% flow)

Analog Output HiRange

(20 mA @ 100% flow)

c. Manual Programming or Reprogramming of the SMM electronics

If electronic parameters were not preprogrammed prior to shipment or
new settings are required, programming of the SMM transmitter with or
without alarms and pulse output may also be done locally via an
Emerson Model 275 hand held Communicator. Refer to Figures 3-1
and 3-2 for Device Description Basic Setup and Detailed Device
Description Setup Trees.generate an electrical signal directly
proportional to the gas mass flow rate.