Electrical installation, Wire the panel, Plumbing – Hach-Lange CLT10 sc User Manual

Page 10: Plumb the panel

Advertising
background image

Electrical installation

Wire the panel

W A R N I N G

Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.

The controller supplies power to the sensors and the digital gateway.

1. Remove power from the controller.
2. Connect the pH sensor (optional) wires to the gateway. For information on how to install sensors

in the flow cells, or how to connect the sensor to the gateway, refer to the sensor user manual.

3. Connect the chlorine sensor to the gateway. For information on how to connect the sensor to the

gateway, refer to the sensor user manual.

Note: When the panel is shipped, the chlorine sensor cable is taped to the front of the panel near the location
of the sensor connection.

4. Connect the controller to the gateway with the 1 m connector cable that is supplied with the

panel.

Plumbing

Plumb the panel

Make sure to follow the size specifications for all panel tubing and fittings. The flow path must
increase in diameter as water flows through the system to prevent build-up of backpressure.

Figure 1

on page 5 shows an overview of the plumbing connections.

1. Attach tubing to the sample inlet and outlet fittings.
2. Make sure the cleaning port on the chlorine flow cell is closed.
3. Make sure the grab sample port on the pH flow cell is closed.
4. Open the flow valve on the sample line and allow water to move through the system.
5. Make sure there are no leaks from the tubing, valve fittings or the flow cells.
6. To minimize air bubbles, the pH flow cell may be tilted between 0º and 30º as shown in

Figure 6

.