Troubleshooting, Event log – Hach-Lange CLT10 sc User Manual

Page 16

Advertising
background image

Troubleshooting

Event log

The controller provides one event log for each sensor. The event log stores a variety of events that
occur on the devices such as calibrations done, calibration options changed, etc. A list of possible
events is shown below. The event log can be read out in a CSV format. For instructions on
downloading the logs, refer to the controller user manual.

Table 1 Event log

Event

Description

Power On

The power was turned on.

Flash Failure

The external flash has failed or is corrupted.

1pointpHCalibration_Start

Start of one-point sample calibration for pH

1pointpHCalibration_End

End of one-point sample calibration for pH

2pointpHCalibration_Start

Start of two-point sample calibration for pH

2pointpHCalibration_End

End of two-point sample calibration for pH

1pointBufferpHCalibration_Start Start of one-point buffer calibration for pH

1pointBufferpHCalibration_End

End of one-point buffer calibration for pH

2pointBufferpHCalibration_Start Start of two-point buffer calibration for pH

2pointBufferpHCalibration_End

End of two-point buffer calibration for pH

TempCalibration_Start

Start of temperature calibration

TempCalibration_End

End of temperature calibration

1pointChemZeroCL2_Start

Start of one-point chemical zero calibration for chlorine

1pointChemZeroCL2_End

End of one-point chemical zero calibration for chlorine

1pointElecZeroCL2_Start

Start of one-point electrical zero calibration for chlorine

1pointElecZeroCL2_End

End of one-point electrical zero calibration for chlorine

1pointProcessConc_Start

Start of one-point process concentration for chlorine

1pointProcessConc _End

End of one-point process concentration for chlorine

2pointChemCL2_Start

Start of two-point chemical calibration for chlorine

2pointChemCL2_End

End of two-point chemical calibration for chlorine

2pointElecCL2_Start

Start of two-point electrical calibration for chlorine

2pointElecCL2_End

End of two-point electrical calibration for chlorine

CL2CalSetDefault

The chlorine calibration was reset to the default.

pHCalSetDefault

The pH calibration data was reset to the default.

TempCalSetDefault

The temperature calibration data was reset to the default.

AllCalSetDefault

All sensor calibration data was reset to the default.

CL2CalOptionChanged

The chlorine calibration option was changed.

pHCalOptionChanged

The pH calibration option was changed.

TempCalOptionChanged

The temperature calibration option was changed.

SensorConfChanged

The sensor configuration was changed.