Table 3 Warning list for the sensor (continued)

Warning

Description

Resolution

PH MAINT RECD

pH sensor maintenance is recommended.
The offset is outside of ±45 mV but within ±60 mV.

Clean the sensor and replace the
salt bridge and standard cell
solution and then repeat the
calibration, or replace the sensor.

T MAINT RECD

The temperature offset is outside of ±3 °C but within
±5 °C.

Calibrate the temperature.

Replacement parts and accessories

W A R N I N G

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Parts

Description

Item no.

Cable, gateway to controller, 1 m

6122400

Cable assembly, 4 pin, shielded, flow sensor, 2 m

9160900

Cable assembly, 5 pin, shielded, M12, chlorine sensor, 1 m

9160300

Fitting, inlet port for chlorine flow cell

English:

9159200 (1/4 in.)

Metric:

9196400 (6 mm)

Fitting, speed fit, outlet port for chlorine flow cell or inlet port for pH flow cell

English:

9159300 (3/8 in.)

Metric:

9196500 (10 mm)

Fitting, elbow, outlet port for pH flow cell

English:

9196900 (1/2 in.)

Metric:

9197000 (12 mm)

Float, flow sensor

9159100

Flow cell assembly, chlorine

English: 9198400

Metric: 9198500

Flow cell assembly, pH combination

English: 9180200

Metric: 9180400

Flow cell assembly, pHD
includes sealing hub

English: 9180100

Metric: 9180300

Flow cell body only (pH)

9159700

Flow cell block, chlorine

9158900

Flow control valve

9159000

Flow control valve kit

9198600

Flow sensor

9160800

English 19

Advertising
This manual is related to the following products: