Relay setup, Relay activation options – Hach-Lange HACH 5500 sc PO43-HR Operations User Manual

Page 12

Advertising
background image

5. Select SET FUNCTION>LOGARITHMIC and then select the

applicable options in the ACTIVATION menu.

Option

Description

SET 50% VALUE

Sets the value corresponding to 50% of the process
variable range.

SET HIGH VALUE Sets the high endpoint (upper value) of the process

variable range.

6. Select SET FUNCTION>BILINEAR and then select the applicable

options in the ACTIVATION menu.

Option

Description

SET LOW VALUE

Sets the low endpoint of the process variable
range.

SET HIGH VALUE

Sets the high endpoint (upper value) of the
process variable range.

SET KNEE POINT
VALUE

Sets the value at which the process variable
range divides into another linear segment.

SET KNEE POINT
CURRENT

Sets the value of the current at the knee point
value.

Relay setup

1. Select SETUP SYSTEM>SETUP OUTPUTS>RELAY SETUP.
2. Select the relay.
3. Select an option.

Option

Description

ACTIVATION

The menu list items change with the selected function.
Refer to

Relay activation options

on page 12 for more

information.

SELECT
SOURCE

Selects the output. Options: None (if the relay is not
configured), the analyzer name or calculation (if a
calculation formula has been configured). Refer to

Set

up the calculation

on page 10.

Option

Description

SET FUNCTION

Selects a function. ALARM—The relay starts when the
upper or lower alarm value is triggered. FEEDER
CONTROL—The relay shows if a process value is larger
or falls below a setpoint. EVENT CONTROL—The relay
toggles if a process value reaches an upper or lower
limit. SCHEDULER—The relay switches at certain times
independently of any process value. WARNING—The
relay shows warning and error conditions in probes.

SET TRANSFER

Selects active or inactive.

FAIL SAFE

Selects yes or no.

Relay activation options

1. Select SETUP SYSTEM>SETUP OUTPUTS>RELAY SETUP.
2. Select the applicable relay.
3. Select ALARM in the FUNCTION menu and then select the

applicable options in the ACTIVATION menu.

Option

Description

LOW ALARM

Sets the value to set the relay to on in response to the
decreasing measured value. For example, if the low alarm
is set for 1.0 and the measured value drops to 0.9, the
relay starts.

HIGH ALARM

Sets the value to set the relay to on in response to the
increasing measured value. For example, if the high alarm
is set for 1.0 and the measured value increases to 1.1, the
relay starts.

LOW
DEADBAND

Sets the range where the relay stays on after the
measured value increases above the low alarm value. For
example, if the low alarm is set for 1.0 and the low
deadband is set for 0.5, the relay stays on between
1.0 and 1.5. The default is 5% of the range.

HIGH
DEADBAND

Sets the range where the relay stays on after the
measured value decreases below the high alarm value.
For example, if the high alarm is set for 4.0 and the high
deadband is set for 0.5, the relay stays on between
3.5 and 4.0. The default is 5% of the range.

12 English

Advertising