Set the error hold mode

1. Select SETUP SYSTEM>SETUP OUTPUTS>ERROR HOLD MODE.
2. Select an option.

Option

Description

HOLD OUTPUTS

Holds the outputs at the last known value when
communications are lost.

TRANSFER
OUTPUTS

Switches to the transfer mode when communications
are lost. The outputs transfer to a pre-defined value.

View data

1. Select VIEW DATA.
2. Select an option.

Option

Description

ANALYZER DATA

Shows specific status information (refer to

Table 2

).

MEASUREMENT
DATA

Shows measurement information (refer to

Table 3

).

LOG DATA

Selects the data log and the event log. DATA LOG
—shows the measurement values. Select start
time, number of hours and number of readings.
EVENT LOG—shows all analyzer information.
Select start time, number of hours and number of
readings.

Table 2 Analyzer data

Element

Definition

CELL TEMP

Temperature of the colorimeter cell block heater (ideally
49.8 °C to 50.2 °C (121.64 °F to 122.36 °F))

REAGENT TEMP

Reagent temperature before it enters the colorimeter

AMBIENT TEMP

Air temperature inside the electronics area

Table 2 Analyzer data (continued)

Element

Definition

SAMPLE TEMP

Sample pre-heater block temperature (typically 45 °C to
55 °C (113 °F to 131 °F) but could reach 58 °C
(136.4 °F))

AIR PRESS

Reagent air pressure in the reagent bottles (ideally
3.95 to 4.10 psi)

LED DUTY CYCLE

Depends on condition of colorimeter cell and age of
analyzer (typically 7,200 to 40,000 counts)

HEATER DUTY
CYCLE

Percentage of time the colorimeter heater is on to
maintain a constant 50 °C (122 °F)

SAMPLE FLOW

Approximate sample flow into the colorimeter,
measured during the flush cycle

SAMPLE PRESS 1

Sample pressure before the pre-heater block (ideally
2 to 4.5 psi depending on incoming sample pressure)

SAMPLE PRESS 2

Differential pressure measured across the sample pre-
heater during a colorimeter cell flush. Slightly less than
SAMPLE PRESS1 during the flush cycle and zero when
the flush is off.

REAGENT 1

Level of reagent remaining

REAGENT 2

Level of reagent remaining

REAGENT 3

Level of reagent remaining

STD SOLUTION

Level of standard solution remaining

FAN SPEED

The speed of the ventilation fan.

LEAK COUNTS

Indication of possible fluid leaks (range 0 to 1023). A
count of over 511 indicates a fluid leak

Table 3 Measurement data

Element

Definition

LAST MEAS CHANNEL Last channel measured.

LAST MEAS TIME

Time of last measurement.