Update the firmware, Calibration, Schedule automatic calibrations – Hach-Lange HACH 5500 sc PO43-HR Operations User Manual

Page 16

Advertising
background image

manufacturer recommends to use an SD card with a minimum of 2 GB
storage capacity.

1. Install the SD card (refer to

Figure 1

on page 4).

2. Select SD CARD SETUP from the MAIN MENU.

Note: The SD CARD SETUP option shows only when an SD card is installed.

3. Select an option.

Option

Description

UPGRADE
SOFTWARE
(conditional)

Shows when an the upgrade file is available on the
SD card. Select the specific device for the
upgrade.

SAVE LOGS

Selects the device for the download data and to
save the logs for last day, last week, last month or
all.

MANAGE
CONFIGURATION

Saves and restores the backup settings, restore
settings and/or transfer settings between
instruments.

WORK WITH DEVICES READ DEVICE FILES—selects the data for each

device to save on the SD card. Options: sensor
diag, measurement data (curve data for a
measurement cycle), cal history, cal data and/or
test script. WRITE DEVICE FILES—shows when
an upgrade file is available for a new measurement
cycle script.

Update the firmware

Use an SD card with an upgrade file to update the firmware for the
controller, sensor or network card. The upgrade menu is shown only
when the SD card contains an upgrade file.

1. Install the SD card into the SD card slot.
2. Select SD CARD SETUP from the MAIN MENU.

Note: The SD CARD SETUP option shows only when an SD card is installed.

3. Select UPGRADE SOFTWARE and confirm. Select the device and

upgrade version, if applicable.

4. When the upgrade is complete, the display shows TRANSFER

COMPLETE. Remove the SD card.

5. Restart the instrument for the upgrade to take effect.

Calibration

The automatic calibration uses the installed known standard to calibrate
the analyzer. Calibration can be done manually or scheduled to occur
automatically. Use the calibration menu to view calibration data, start or
cancel the auto calibration, override the auto cal settings, calibrate
manually or to reset to the default calibration.

1. Push cal to display the calibration menu.

Option

Description

START MANUAL CAL

Refer to

Start a calibration manually

on page 17.

SET AUTO
CALIBRATION

Refer to

Schedule automatic calibrations

on page 16.

CALIBRATION DATA

Displays data from the last calibration and the
due date and time of the next scheduled
calibration.

OVERRIDE
CALIBRATION

Enter new slope and zero values and disables
automatic calibration.

OUTPUT CALIBRATION Select a 4-20 mA output and enter the output

values to be sent.

RESET DEFAULT CAL

Reset calibration data to default values and
disables automatic calibration. On completion,
perform a new calibration.

Schedule automatic calibrations

1. Push cal, then select SET AUTO CALIBRATION.
2. Select ACTIVATE AUTO CAL>YES.

16 English