Start a calibration manually – Hach-Lange HACH 5500 sc PO43-HR Operations User Manual

Page 17

Advertising
background image

3. Select STD SOLUTION and enter the standard value in ppb.
4. Select a schedule option for calibration.

Option

Description

TIME BASE

Sets the interval between calibrations. Options: DAYS or
HOURS.

WEEK DAY

Selects the day or days of the week for calibration when
TIME BASE is set to DAY.

TIME

Sets the time of day for calibration when the TIME BASE is
set to DAY.

SET INTERVAL Sets the interval between automatic calibrations in hours

when the TIME BASE is set to HOURS.

Start a calibration manually

1. Push cal, then select START MANUAL CAL.
2. The type of calibration is displayed. Push enter to start the

calibration.

3. The measurement status is displayed. Select YES to interrupt the

current measurement cycle and start the calibration immediately.
Select NO to wait until the current measurement cycle has completed
before the calibration starts.

English 17