Option

Description

OFF DELAY

Sets a delay time (0–300 seconds) to set the relay off
(default = 5 seconds).

ON DELAY

Sets a delay time (0–300 seconds) to set the relay on
(default = 5 seconds).

4. Select FEEDER CONTROL in the FUNCTION menu and then select

the applicable options in the ACTIVATION menu.

Option

Description

PHASE

Specifies the relay status if the process value is larger
than the setpoint. HIGH(default)—sets the relay to on
when the process value is larger than the setpoint. LOW
—sets the relay to on when the process value falls below
the setpoint.

SET SETPOINT

Sets the process value for the relay to toggle between the
high and low value (default = 10).

DEADBAND

Sets a delay so that the relay will be stable as the
process value converges to the setpoint.

OVERFEED
TIMER

Sets the maximum amount of time to get to the process
setpoint. When the time expires and the relay does not
show the setpoint, the relay is set to off. After an overfeed
alarm occurs, reset the timer manually.

OFF DELAY

Sets a delay time to set the relay to off (default =
5 seconds).

ON DELAY

Sets a delay time to set the relay to on (default =
5 seconds).

5. Select EVENT CONTROL in the FUNCTION menu and then select

the applicable options in the ACTIVATION menu.

Option

Description

SET SETPOINT Sets the value to set the relay to on.

DEADBAND

Sets a delay so that the relay will be stable as the process
value converges to the setpoint.

OnMax TIMER

Sets the maximum time the relay stays on (default =
0 min).

Option

Description

OffMax TIMER

Sets the maximum time the relay stays off (default =
0 min).

OnMin TIMER

Sets the time the relay stays on, independent from the
measured value (default = 0 min).

OffMin TIMER

Sets the time the relay stays off, independent from the
measured value (default = 0 min).

6. Select SCHEDULER in the FUNCTION menu and then select the

applicable options in the ACTIVATION menu.

Option

Description

HOLD OUTPUTS Holds or transfers outputs for the selected channels.

RUN DAYS

Selects the days the relay stays on. Options: Sunday,
Monday, Tuesday, Wednesday, Thursday, Friday,
Saturday

START TIME

Sets the start time.

INTERVAL

Sets the time between activation cycles (default = 5 min).

DURATION

Sets the period of time the relay is set to on (default =
30 sec).

OFF DELAY

Sets the time for additional hold/output time after the
relay is set to off.

7. Select WARNING in the FUNCTION menu and then select the

applicable options in the ACTIVATION menu.

Option

Description

WARNING LEVEL Sets the level for the warning activation and starts

applicable individual warnings.