Display values, Process data controller block – Hach-Lange SC 200 Profibus network card User Manual User Manual

Page 10

Advertising
background image

When the manual configuration is selected, the telegram data structure
can be configured by the user (refer to

Setup the network

on page 8).

Table 2 Profibus data telegram structure

Byte number

Data

Data type

1–2

Classified error

Integer (2 bytes)

3–4

Classified status

Integer (2 bytes)

5–8

Measurement 1

Floating (4 bytes)

9–12

Measurement 2

Floating (4 bytes)

13–16

Measurement 3

Floating (4 bytes)

Display values
The Profibus data block structure (

Profibus message data block

structure

) can replace sc probes without changes in the PLC configuration.

The primary value is always the measured value.
The secondary value, if not available, is filled with zero.
The tertiary value, if not available, is filled with zero.

Figure 2 Profibus message data block structure

Process data controller block
The data block for the sc controller is similar to the data block for sensors.
The structure of the sc controller data block is independent of the number
of connected sensors:

• sc controller_ERROR
• sc controller_STATUS
• Primary value
• Secondary value
• Tertiary value

Block 3 sc controller ERROR

and

Block 3 sc controller STATUS

show the

data definitions for error and status 1 in the sc controller.

Table 3 Block 3 sc controller ERROR

Bit

Error

Note

0

Sensor 1 communication
error

A communication error has occurred
between the sc controller and sensor 1, the
sensor could be disconnected.

1

Sensor 2 communication
error

A communication error has occurred
between the sc controller and sensor 2, the
sensor could be disconnected.

2–15 Not used

Table 4 Block 3 sc controller STATUS

Bit

Status1

Note

0

Sensor 1 installed First sensor has been installed to the sc controller. This

bit is set even if the sensor is disconnected after
installation.

1

Sensor 2 installed Second sensor has been installed to the sc controller.

This bit is set even if the sensor is disconnected after
installation.

2

Relay A on

3

Relay B on

10 English