Troubleshooting, Error and status indicators – Hach-Lange SC 200 Profibus network card User Manual User Manual

Page 12

Advertising
background image

Troubleshooting

W A R N I N G

Multiple hazards. Do not disassemble the instrument for maintenance or service.
If the internal components must be cleaned or repaired, contact the manufacturer.

Error and status indicators

Error and status words follow the same standard definition for all sc probes
and controllers.

Error messages

lists bit position and error messages.

Status indicator

messages

lists bit position and status messages.

A bit value of zero shows the error or status condition that is not true.
A bit value of 1 shows the error or status condition that is true. For example,
if Bit 0 has the value of 1, an error has occurred during the last calibration.

Table 6 Error messages

Bit Message

Indication

0

Measurement calibration
error

An error has occurred during the last calibration

1

Electronic adjustment error An error has occurred during the last electronic

calibration

2

Cleaning error

The last cleaning cycle failed

3

Measuring module error

A failure has been detected in the Measurement
Module

4

System re-initialization
error

Some settings are inconsistent and have been
reset to factory defaults

5

Hardware error

A general hardware error has been detected

6

Internal communication
error

A communication failure within the device has
been detected

7

Humidity error

Excessive humidity has been detected within
the device

8

Temperature error

Temperature within the device exceeds a
specified limit

Table 6 Error messages (continued)

Bit Message

Indication

9

10

Sample warning

Some action is required with the sample system

11

Questionable calibration
warning

The last calibration may not be accurate

12

Questionable
measurement warning

One or more of the device measurements are
out of range or are of questionable accuracy

13

Safety warning

A condition has been detected which may result
in a safety hazard

14

Reagent warning

The reagent system requires attention

15

Maintenance required
warning

The device requires maintenance

Table 7 Status indicator messages

Bit Message

Indication

0

Calibration in progress

The device is in a calibration mode.
Measurements may not be valid.

1

Cleaning in progress

The device is in a cleaning mode.
Measurements may not be valid.

2

Service/Maintenance menu

The device is in a service or maintenance
mode. Measurements may not be valid.

3

Common error

The device has recognized an error. See Error
Register for Error Class.

4

Measurement 0 Quality Bad Precision of measurement is out of specified

limits.

5

Measurement 0 Low Limit

Measurement is below the specified range.

6

Measurement 0 High Limit

Measurement is above the specified range.

7

Measurement 1 Quality Bad Precision of measurement is out of the

specified limits.

12 English