2 data table – Hach-Lange LCK 348_349_350 User Manual
Page 22
Advertising
22
•
LCK 348: 0.5-5 mg/l PO
4
-P or 1.5-15 mg/l PO
4
•
LCK 349: 0.05-1.5 mg/l PO
4
-P or 0.15-4.5 mg/l PO
4
•
LCK 350: 2-20 mg/l PO
4
-P or 6-60 mg/l PO
4
Select the calibration data of maximum two measuring ranges which will be the
suitable one.
Each programming procedure has only to be followed once prior to the first test
evaluation.
For detailed information please read the POCKET COLORIMETER II instruction
manual
.
3.2.2 Data table
3.2.2.1
Evaluation as Phosphate-Phosphorus PO
4
-P
Table 1 Evaluation as Phosphate-Phosphorus PO
4
-P
LCK 348
LCK 349
LCK 350
RES (resolution)
--.--
--.--
--.--
S0 (standard 0)
00.50
00.05
02.00
A0 (absorbance 0)
0.104
0.090
0.104
S1 (standard 1)
05.00
01.50
20.00
A1 (absorbance 1)
0.600
0.728
0.753