2 data table – Hach-Lange LCK 348_349_350 User Manual

Page 22

Advertising
background image

22

LCK 348: 0.5-5 mg/l PO

4

-P or 1.5-15 mg/l PO

4

LCK 349: 0.05-1.5 mg/l PO

4

-P or 0.15-4.5 mg/l PO

4

LCK 350: 2-20 mg/l PO

4

-P or 6-60 mg/l PO

4

Select the calibration data of maximum two measuring ranges which will be the

suitable one.

Each programming procedure has only to be followed once prior to the first test

evaluation.

For detailed information please read the POCKET COLORIMETER II instruction
manual

.

3.2.2 Data table

3.2.2.1

Evaluation as Phosphate-Phosphorus PO

4

-P

Table 1 Evaluation as Phosphate-Phosphorus PO

4

-P

LCK 348

LCK 349

LCK 350

RES (resolution)

--.--

--.--

--.--

S0 (standard 0)

00.50

00.05

02.00

A0 (absorbance 0)

0.104

0.090

0.104

S1 (standard 1)

05.00

01.50

20.00

A1 (absorbance 1)

0.600

0.728

0.753