3 programming procedure, 1 lck 348 phosphate-phosphorus po – Hach-Lange LCK 348_349_350 User Manual

Page 23

Advertising
background image

23

3.2.2.2

Evaluation as Phosphate PO

4

3.3 Programming procedure

Note: Each programming procedure has only to be followed once prior to the first test
evaluation.

Important Note: The following programming procedures are valid for the evaluation
form Phosphate-Phosphorus PO

4

-P. To evaluate as Phosphate PO

4

please choose

the data mentioned in table

3.2.2.2 Evaluation as Phosphate PO4 on page 23

.

Note: After programming the calibration data, it could be helpful to mark the channel(s) 1
and/or 2 accordingly with a label.

3.3.1 LCK 348 Phosphate-Phosphorus PO

4

-P

1.

Press POWER to switch on the instrument.

2.

Press MENU. "SEL" is displayed. A flashing arrow indicates the current

measuring range (1 or 2).

3.

Select measuring range 1 or 2 for LCK 348 (0.5-5 mg/l PO

4

-P).

Note: Press

GREEN

key to toggle between measuring ranges.

4.

Press MENU for about five seconds until "USER" followed by "CAL" is

displayed.

5.

Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is

displayed.

6.

Press GREEN key. "-.---" is displayed.

7.

Press BLUE key until "--.--" is displayed. Press GREEN key. "S0" is displayed.

8.

Press GREEN key until "A0" is displayed.

9.

Press GREEN key until "S1" is displayed.

10. Press GREEN key. Enter "05.00" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A1" is

displayed.

11. Press GREEN key. Enter "0.600" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "Add" is

displayed.

Table 2 Evaluation as Phosphate PO

4

LCK 348

LCK 349

LCK 350

RES (resolution)

--.--

--.--

--.--

S0 (standard 0)

01.50

00.15

06.00

A0 (absorbance 0)

0.104

0.090

0.104

S1 (standard 1)

15.00

04.50

60.00

A1 (absorbance 1)

0.600

0.728

0.753