5 overwrite stored calibration data, Rst. see, 1 retrieve the factory – Hach-Lange LCK 348_349_350 User Manual

Page 27: 1 retrieve the factory calibration

Advertising
background image

27

3.

"RES" is displayed. Press BLUE key to scroll through the entries. Compare the

entries with the figures in the

3.2.2.1 Evaluation as Phosphate-Phosphorus

PO4-P on page 22

.

In case a value was not stored correctly, retrieve the factory calibration first. See

3.5.1

Retrieve the factory calibration on page 27

. After retrieving the factory calibration,

start the programming procedure again.

3.5 Overwrite stored calibration data

With the POCKET COLORIMETER II it is possible to overwrite stored calibration

data. Before overwriting calibration data it is recommended to retrieve the factory

calibration first. After retrieving the factory calibration, you can program new

calibration data.

In case of any mistake retrieve the factory calibration first.

3.5.1 Retrieve the factory calibration

1.

Press MENU. "SEL" is displayed. A flashing arrow indicates the current

measuring range (1 or 2). Press GREEN key to toggle between measuring

ranges. Select measuring range retrieving the factory calibration.

1.

Press MENU for about five seconds until "USER" followed by "CAL" is

displayed.

2.

Press BLUE key to scroll to "dFL". Press GREEN key.

Advertising