3 lck 350 phosphate-phosphorus po – Hach-Lange LCK 348_349_350 User Manual

Page 25

Advertising
background image

25

10. Press GREEN key. Enter "01.50" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A1" is

displayed.

11. Press GREEN key. Enter "0.728" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "Add" is

displayed.

12. Press MENU until "Edit" is displayed. Press GREEN key until „S0“ is displayed.
13. Press GREEN key. Enter "00.05" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until "A0" is

displayed.

14. Press GREEN key. Enter "0.090" by using BLUE key to change value and

GREEN key to move to the next space. Press GREEN key until „S1“ is

displayed.

15. Press MENU until "Edit" is displayed to check the entered data. Press GREEN

key. „RES“ is displayed.

16. Press BLUE key to scroll through the entries. Compare the entries with the data

shown in the table below.

17. After comparing the entries press MENU two times to enter the measuring

mode. See

4.3 Evaluate LANGE Phosphate-Phosphorus/ Phosphate Cuvette

Tests on page 29

.

Note: In case of any mistake retrieve the factory calibration first. See

3.5.1 Retrieve the factory

calibration on page 27

. After retrieving the factory calibration, start the programming procedure

again. If the instrument is shut-off or loses power during data entry, all edits will be lost.
Automatic shut-off is 60 minutes.

3.3.3 LCK 350 Phosphate-Phosphorus PO

4

-P

1.

Press POWER to switch on the instrument.

2.

Press MENU. "SEL" is displayed. A flashing arrow indicates the current

measuring range (1 or 2).

3.

Select measuring range 1 or 2 for LCK 350 (2-20 mg/l PO

4

-P).

Note: Press

GREEN

key to toggle between measuring ranges.

4.

Press MENU for about five seconds until "USER" followed by "CAL" is

displayed.

5.

Press BLUE key until "Edit" is displayed. Press GREEN key. " RES" is

displayed.

6.

Press GREEN key. "-.---" is displayed.

Table 4 Calibration data of LCK 349

RES

--.--

S0

0.05

A0

0.090

S1

1.50

A1

0.728