Installation – Hach-Lange ORBISPHERE 3650 User Manual

Page 24

Advertising
background image

22

Installation

The interface box operates from 115 VAC or 230 VAC power. Make sure that the power is
correct before connecting to a power supply. A green power LED is illuminated when the box is
plugged into the power source.

Two cables are supplied with the model 29122 interface box:

A model 32511 cable for connection between the 3650EX instrument LEMO-6 connector and
the interface box

A model 32538 cable for connection between the interface box and the PC.

The connections should be made as indicated in

Figure 3

.

Note:

When the Interface Box is connected to the 3650Ex instrument, this also acts as a power supply to

the instrument and disconnects the internal battery power source.

Note:

If you use an adapter for the connection to the PC, make sure it is designed for this purpose and,

thus, has all nine pins accessible. Some 25-to-9 pin adapters are supplied for specific use, such as a
mouse, and these may have only certain pins available.

Pseudo RS232 Signal

LEMO-6

Pin

Transmitted data (TTL-TXD)

Pin 1

Received data (TTL-RXD)

Pin 2

Not used

Pin 3

Not used

Pin 4

External voltage input (used with the model 29122 interface,
only

in safe area)

Pin 5

Ground

Pin 6

Pin

Use

Pin 1 Not Used

Pin 2 TTL Transmitted Data (TXD)

Pin 3 TTL Received Data (RXD)

Pin 4 Not Used

Pin 5 Power Supply (V+)

Pin 6 Not Used

Pin 7 Not Used

Pin 8 Not Used

Pin 9 Ground

Pin

Use

Pin 1 Not Used

Pin 2 RS-232 Transmitted Data (TXD)

Pin 3 RS-232 Received Data (RXD)

Pin 4 Not Used

Pin 5 Ground

Pin 6 Not Used

Pin 7 Not Used

Pin 8 Not Used

Pin 9 Not Used