Installation

3.6 Installation completion check list

3.6.1 Battery

The instrument is designed to work on battery power.

Install the Exproof lithium battery by first unscrewing the instrument's battery cap located on the
right side of the instrument (see

Figure 1 on page 19

) with a coin or flat screwdriver. Then insert

the battery pack lengthwise, positive end first, and replace the cap.

The power autonomy of the instrument is about 60 hours with a new battery. If battery power
should drop, a LO BAT warning appears in the instrument LCD's top-left corner.

3.6.2 Electrical connections

The LEMO-6 connector on the right side of the instrument (see

Figure 1 on page 19

) is used for

the connection to a PC. To connect the instrument to a PC, use the model 29122 interface box
(see also

3650Ex Instrument - PC connection on page 21

).

Remember, connecting the instrument to the interface box means it is now powered by mains
power and the battery power is by-passed.

3.6.3 Instrument clock setting

If you use the instrument to store measurements for downloading to a PC, you should verify the
date and time settings of the instrument's internal clock, as described in

Clock settings on

page 46

.

3.6.4 Electrochemical sensor

Before making any measurements, for A1100 ATEX sensors perform the preparation procedure
in the A1100 Sensor Manual. For other ATEX sensors perform the sensor service procedure as
described in the EC Sensor Manual.

3.6.5 Flow chamber

The model 32007E flow chamber's inlet and outlet should be free of any obstructions. It is
mounted by means of the sensor's collar, as shown in

Figure 2 on page 20

. A guide pin on the

flow chamber surface prevents twisting during operation.

When switching from liquid to gaseous samples, ensure that the sensor membrane is dry.
Ensure that the gas from the exit tube of the flow chamber (in gaseous mode) is released at
atmospheric pressure and that it is constant.

3.7 Storage when not used

At the end of the workday clean the outside of the instrument and interface box (if used) with a
damp cloth. Run clean water through the flow chamber to prevent passageways from clogging.
You may wish to repeat sensor preconditioning (see

Preconditioning sensors on page 27

) prior

to the next use.

If you do not expect not to use your sensor for more than a few months, you should clean the
sensor as instructed in the Sensor Manual and then store it dry and with the calibration cap in
place for protection.

Only the model 32960 battery can be used with this instrument.

The battery may be installed or changed in the hazardous area

The instrument to PC connection should be used only in a safe area.