4 intrinsically safe conformity, General information – Hach-Lange ORBISPHERE 3650 User Manual

Page 8

Advertising
background image

6

General Information

1.4 Intrinsically safe conformity

Orbisphere series 3650Ex analyzers for gas measurement have been certified as Intrinsically
Safe by:

LCIE (Laboratoire Central des Industries Electriques),
33 av. Division Leclerc, Fontenay aux Roses 92260, France.

Note:

LCIE is a notified body number 0081 in accordance with article 9 of Directive 94/9/CE of the

European Parliament and Council of 23 March 1994.

LCIE certifies that the electrical apparatus has been found to comply with the essential Health
and Safety Requirements: EN 60079-0 (2006), EN 60079-11 (2007).

These instruments are certified

II 1G Ex ia IIC T4 under EC type Examination Certificate

number LCIE 03 ATEX 6003 X (see

3650Ex certificates on page 13

)

Category

Explanation

II 1 G

ATEX marking: Surface apparatus with permanent explosive gas presence.

Ex

Explosion-proof apparatus built to the universal standards below:

ia

Type of protection: The highest category, based on a safety factor of 1.5 on two faults.
No combination of two faults in the 3650Ex can produce a spark, or heating, causing
ignition of an explosive atmosphere.

IIC

Gas group: Corresponds to the most flammable gases, including hydrogen.

T4

Temperature category: Maximum surface temperature of 135º C.

Advertising