Connector J8 (sensor)
The numbers listed below refer to the 10 available J8 connections (from left to right) in

Figure 8

. The

colors indicated are the wire colors in the sensor cable.

Note: Remember, this connector is colored orange for EC sensors and black for TC sensors.

A1100 EC sensor

31xxx EC sensor

31xxxS smart EC
sensor

TC sensor

Sensor
cable
wall or
panel

LEMO
10 cable
portable

1.

Guard electrode

Guard electrode

Guard electrode

GND for power

Yellow

Brown

2.

RS485A+

Not used

I2C-SCL

V2 signal

Pink

Red

3.

Thermistor A

Thermistor A

Thermistor A

Solenoid

Grey

Orange

4.

Anode electrode

Anode electrode

Anode electrode

Relay coil

Red

Yellow

5.

RS485B

Not used

I2C-SDA

+12V power

Purple

Green

6.

Thermistor B

Thermistor B

Thermistor B

+24V power

White

Blue

7.

GND

Not used

GND

V3 signal

Black

Purple

8.

+ 5V

Not used

+ 5V

GND for signal

Green

Grey

9.

Cathode electrode

Cathode electrode

Cathode electrode

-5V power

Blue

White

10.

Not used

Not used

Not used

Temperature

Brown

Black

Measurement alarm relays

The three output relays are located on the measurement board.
They can be individually configured to Normally Open (NO) or to Normally Closed (NC) by
physically moving the jumper on each relay. The illustration is for the EC measurement
board (the positions are different for the TC measurement board):

• Upper relay is set to NC
• Middle relay is set to NO
• Lower relay is shown with no jumper

Note: For all measurement boards, J4 is relay 1, J5 is relay 2 and J6 is relay 3

Sensor installation

EC sensors

For EC sensor installation, servicing, and maintenance ensure you follow the instructions in the
Sensor Installation and Maintenance manual that was supplied with the instrument.

TC sensors

For TC sensor installation, servicing and maintenance ensure you follow the instructions in the TC
Sensor Installation and Maintenance
manual that was supplied with the instrument. Pay particular
attention to the installation and connection of the purge gas supply.

N O T I C E

Do not place the TC sensor into a liquid sample until a constant supply of dry purge gas has been connected, as
liquid could condense inside the measuring chamber and cause damage to the thermal conductor chip.

To ensure the continuation of purge gas while the sensor is in contact with the sample, it is highly
recommended to use a backup purge gas cylinder with an automatic changeover valve that activates
when the first cylinder is empty.
The use of an ORBISPHERE Model 29089 gas regulator (or similar) is also recommended to deliver
a constant, pressure regulated supply of dry purge gas to the sensor, filtered to 40 μm.