Basic user manual, Specifications, English – Hach-Lange ORBISPHERE 510_51x Basic User Manual User Manual

Page 3

Advertising
background image

Specifications

Specifications are subject to change without notice.

Specification

Details

Operating temperature

-5°C to +40°C

Storage temperature

-20°C to +70°C

Operating humidity

0 to 95% non condensing relative humidity

Operating altitude

From 0 to 2,000 m. (6,550 ft.) above sea level

EMC requirements

EN61326-1:2006 Directive 2004/108/EEC

Note: The wall mount instrument is a Class A product. In a domestic environment this
product may cause radio interference in which case the user may be required to take
adequate measures.

CE compliance

EN61010-1: 2010 Directive 2006/95/EEC

Safety rating

ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1

Enclosure ratings

IP 65; Totally protected against dust; Protected against low pressure jets of
water from all directions.

Note: For portable models, IP 65 is applicable only when a cable is connected to the
ethernet socket.

NEMA 4X (wall mount only); Totally protected against dust; Protected against
pressure jets of water from all directions.

Power supply
Wall and panel mount

Universal 100 VAC to 240 VAC @ 50/60 Hz - 40VA 10 to 30 VDC - 30W

Power supply
Portable

5V supplied through an external power supply 100-240 VAC @ 50/60 Hz - 1A
Optional battery pack with 4 hour autonomy (charges in about 6 hours). Battery
charge time can increase at T > 30°C.

Analog current output version
on the measurement board(s)

4-20 mA (default) or 0-20 mA (configuration with software);
3 configurable outputs: Maximum load: 500 ohm; Sensitivity: 20μA;
Accuracy: ± 0.5% (between operating temperature limits)

Analog voltage output version
on the measurement board(s)

0- 5 V output (hardware option);
3 configurable outputs; Minimum load: 10 KOhm; Sensitivity: 5 mV;
Accuracy: ± 0.5% (between operating temperature limits)

Measurement alarm relays on
the measurement board(s)

Three alarm relays per measurement board; 1A-30 VAC or 0.5A-50 VDC on a
resistance load.
Configurable to Normally Open [NO] or Normally Closed [NC] contacts by
changing the jumper positions.

W A R N I N G

Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS

System alarm relay on the main
board

One system alarm relay; 1A-30 VAC or 0.5A-50 VDC on a resistance load.
Normally closed [NC] (NO relay also available) when instrument is turned on.

W A R N I N G

Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS

Options

RS-485 or PROFIBUS-DP; USB client; USB host; Ethernet 10/100 Base-T

Wall and pipe mount instrument
(H x D x W)

236.5 x 160 x 250 mm - weight 3.8 kg
9.31 x 6.30 x 9.84 ins. - weight 8.38 lbs