Calibration, Portable calibration device – Hach-Lange ORBISPHERE KM1100 Basic User Manual User Manual

Page 18

Advertising
background image

Note: Data stored in volatile memory is lost when instrument is powered off; non-volatile memory is permanent. In
case of an accidental power off event the instrument resumes measurement storage after the last measurement
stored in flash.

1. Select Measurement file from the Measurement menu:

Option

Description

Storage mode

Select No storage if data storage is not required. Select Store once to start
recording measurements. When the volatile memory is full the recording of
measurement stops. Select Rolling buffer for continuous recording of
measurements. When the volatile memory is full the latest measurement set
replaces the oldest one (first-in, first-out)

RAM time

The delay between two recordings of measurement data.

FLASH time

The delay between two data file transfers from volatile memory into non-volatile
memory. The last data file erases the previous one. This option is only available if
the Auto save in flash box is checked.

Save in flash now

Press this button to store measurement data in flash immediately. After pressing
this button, press OK to initiate the process. A warning screen appears informing
you that the operation can take up to 30 seconds. Press Yes to continue or No to
abort.

Auto save in flash

Check this box to save measurements in flash automatically. Measurements are
saved at regular time intervals as defined in the FLASH time box.

Purge data

Clear all data in the volatile and non-volatile memories.

Start logging
measurements

Only available in Store once mode, this option starts or stops the measurement
recording session. Measurement recording ends automatically when the buffer is
full.

Open data

Opens a table showing the measurments stored in the volatile memory (RAM).

Calibration

Calibrations can only be performed once the instrument has been installed and configured.

Note: The temperature sensor is factory calibrated and can only be changed by a Hach Lange representative.

Portable calibration device

The portable calibration device (part number 33088)
is ideal for calibrating the sensor close to its location
in the sample. The device is designed to hold a one
liter gas cylinder but is fully compatible for use with
other kinds of gas cylinders.
When using other types of gas cylinders, simply
connect the gas supply to the Swagelok

connector

(No. 1), and make sure that the inlet gas pressure
does not exceed 2 bars absolute.
The gas bottle is not supplied and must be
purchased locally.

To ensure the calibration works correctly the calibration gas bottles must be of 99.999% (50) quality
or better. Bottles containing 34 liters of compressed gas with a 5/8-18 UNF (C10) fitting are
compatible with the hand-held calibration device and recommended for this purpose. Empty gas
bottles are easily removed from the device. Simply unscrew the small screw (No. 2) a few turns, slide

18 English