General information, Safety information – Hach-Lange ORBISPHERE KM1100 Basic User Manual User Manual

Page 4

Advertising
background image

Specification

Details

Limit of detection (LOD)

Low range sensors: 0.6 ppb
High range sensors: 0.015 ppm

Response time (90%)

Low range sensors: < 10 seconds in gas phase; < 30 seconds in liquid
phase
High range sensors: < 10 seconds in gas phase; < 50 seconds in liquid
phase

Display resolution

0.1 ppb

Calibration

Low range sensors: Single point calibration (zero)
High range sensors: Two at cap replacement (zero and air), one during use
(air)

Calibration sample

Low range sensors: Standard 99.999% N

2

(quality 50) or equivalent oxygen

free gas
High range sensors: Standard 99.999% N

2

(quality 30) or equivalent oxygen

free gas, air

Ambient temperature

-5 to 50°C (23 to 122°F)

Humidity

0 to 95% non-condensing relative humidity

Power supply

Universal 85-264 VAC @ 50/60 Hz, 25 VA

10-36 VDC, 25 W

Wall and pipe mount instrument
(H x D x W)

236.5 x 160 x 250 mm - weight 3.8 kg
9.31 x 6.30 x 9.84 ins. - weight 8.38 lbs

Panel mount instrument
(housing) (H x D x W)

156 (123) x 250 x 220 (214) mm - weight 2.9 kg
6.14 (4.84) x 9.84 x 8.86 (8.43) ins. - weight 6.39 lbs

M1100 12 mm (PG 13.5) sensor
(L x W)

246 x 47 mm - weight 0.6 kg
9.69 x 1.85 ins - weight 1.32 lbs

K1100 and M1100 28 mm sensor
(L x W)

143.50 x 49 mm - weight 0.74 kg
5.65 x 1.93 ins - weight 1.63 lbs

Calibration device

Weight 0.7 kg

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer’s website.

Safety information

N O T I C E

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.

Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.

4 English