Specifications

Specifications are subject to change without notice.

Specification

Details

EMC requirements

EN61326-1:2006 Directive 2004/108/EEC

Note: The wall mount instrument is a Class A product. In a domestic environment this
product may cause radio interference in which case the user may be required to take
adequate measures.

CE compliance

EN61010-1: 2010 Directive 2006/95/EEC

Safety rating

ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1

Enclosure ratings

IP 65; Totally protected against dust; Protected against low pressure jets of
water from all directions.
NEMA 4X (wall mount only); Totally protected against dust; Protected
against pressure jets of water from all directions.

Analog outputs

3 Smart 0/4 – 20 mA (500 Ohms), programmable as linear or tri-linear,
configurable to send diagnostics or alarm information.

Measurement alarm relays on the
measurement board

Three alarm relays; 1A-30 VAC or 0.5A-50 VDC on a resistance load
Configurable to Normally Open [NO] or Normally Closed [NC] contacts by
changing the jumper positions

W A R N I N G

Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS

System alarm relay on the main
board

One system alarm relay; 1A-30 VAC or 0.5A-50 VDC on a resistance load
Normally closed [NC] (NO relay also available) when instrument is turned on

W A R N I N G

Potential Electrocution Hazard. Connect only safety low voltage < 33 VAC
RMS

Digital communication

RS485, Profibus DP (optional), Ethernet, USB-client to download data from
a computer, USB-host to download data with a USB memory stick

Data storage

Rolling buffer or store once mode for up to 1,000 measurements

Holds calibration records for the last 10 calibrations

Sample temperature

Measurement from -5 to 50°C (23 to 122°F)

Sensor resistant to temperature from -5 to 100°C (23 to 212°F)

Sample pressure

1 to 20 bar absolute (14.5 to 290 psia)

Sample types

K1100 sensor: Water only
M1100 sensor (low range): Water and beer
M1100 sensor (high range): Water, beer, wine, wort and carbonated drinks

Measurement range

Low range sensors: 0 to 2,000 ppb (dissolved)
High range sensors: 0 to 40 ppm (dissolved)

Repeatability

Low range sensors: ± 0.4 ppb or 1%, whichever is the greater
High range sensors: ± 0.015 ppm or ± 2% whichever is the greater

Reproducibility

Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater

Accuracy

Low range sensors: ± 0.8 ppb or 2%, whichever is the greater
High range sensors: ± 0.02 ppm or ± 3% whichever is the greater