Mains connection – Hach-Lange POLYMETRON 9526 User Manual

Page 12

Advertising
background image

Figure 3 Installation off-line

Mains connection

W A R N I N G

The installation of the instrument should be performed exclusively by personnel specialized and authorized to
work on electrical installations, in accordance with relevant local regulations. In addition, and in accordance with
safety standards, it must be possible to disconnect the power supply of the instrument in its immediate vicinity.

Use a three wire mains supply cable (live, neutral and earth) with a cross-section between 0.35 and
2 mm

2

(AWG 22 to 14) rated at 105 °C minimum. The external cable insulation should be cut as

close as possible to the terminal block.
The connector for the mains power cable is delivered with the instrument (see

Figure 4

) and can be

found in the accessory drawer (No. 7 in

Figure 1

on page 8) at the front of the instrument.

Figure 4 Mains power cable connector

1 Live wire

5 Female contact with locking nut

9 Rubber gasket

2 Neutral wire

6 Rubber gasket

10 Cable tightening nut

3 Not used

7 Main connector body

4 Earth wire

8 Clamping ring

12 English