Hach-Lange POLYMETRON 9526 User Manual

Page 3

Advertising
background image

Table of contents

Specifications

on page 3

Startup

on page 13

General information

on page 5

Maintenance

on page 18

Installation

on page 9

Troubleshooting

on page 19

User interface and navigation

on page 13

Additional information

Additional information is available on the manufacturer's website.

Specifications

Specifications are subject to change without notice.

Analyzer

Specification

Details

Dimensions

Height: 450 mm; Width: 250 mm; Depth: 460 mm

Weight

7 kg (15.4 lb)

Casing protection

IP 65 / NEMA4X

Power supply

Standard version: 100-240 VAC 50/60 Hz

Low voltage version: 13-30 VAC 50/60 Hz, 18-42 VDC

Consumption: 25 VA

Measurement category: I (overvoltage less than 1,500 V)

Sample flow rate

20 liters/hour minimum

Sample tubing

Sample inlet and outlet: Diameter 8 mm (or 5/16'') semi-rigid tubing. We recommend the
use of PE tubing if sample temperature is inferior to 70 °C, and PTFE if superior to 70 °C

Connections

Power supply: Use the connector provided in the drawer

Analog output: Use the recommended POLYMETRON cable

Ambient temperature

-20 to 60 °C (-4 to 140 °F)

Maximum temperature 100 °C (at atmospheric pressure)

Maximum pressure

10 bar at ambient temperature

Relative humidity

10—90%

Precision

Conductivity: ± 2% of the displayed value

Temperature: ± 0.2 °C

Measurement range

Conductivity: 0.01 μS/cm to 200 μS/cm

Resistivity: 100 MΩ.cm to 5 kΩ.cm

Temperature: -20 to 200 °C (-4 to 392 °F)

Display resolution

0.001 μS/cm or 0.1 MΩ.cm

English 3