Calibration, About sensor calibration, Calibration process – Hach-Lange POLYMETRON 9526 User Manual

Page 17

Advertising
background image

Calibration

About sensor calibration

There are no calibration options available from the menus of the 9526 instrument. All calibrations are
performed from the controller of the sensor under test. For detailed information on these calibration
procedures refer to the associated manuals delivered with the sensor and controller under test.

N O T I C E

After turning on the 9526 instrument and initiating sample circulation, wait for at least 30 minutes in order to allow
correct rinsing of the whole system. This also enables the temperature equilibration between the sample, the flow
chamber and the sensor.

After 30 minutes of sample circulation, compare the measurement value displayed on the controller
under test against the measurement value displayed on the 9526 instrument. If these values are
outside ± 5% of each other than a calibration is required. If these values are within ± 5% of each
other than a calibration is not necessary but can still be performed.
Before calibrating the sensor under test, make sure to perform a temperature calibration first.

Calibration process

All calibrations are made using the controller and sensor under test. Follow the instructions in the
associated controller and sensor user manuals.
The process can be different according to the Polymetron controller under test. Perform the
calibration process in the following sequence.
1. Temperature calibration
The following equipment is required for a temperature calibration:

• Pt100 simulator (< 0.1 °C) for a 2-point electrical calibration
• Certified precision thermometer (< 0.1 °C) if connected in-line
• None if mounted off-line as the 9526 is used as the reference

Polymetron controller model under test

9500

9125

Others

2-point electrical
calibration

NO

YES (100 and 172 Ω)

NO

Process calibration

YES

YES

YES

2. Electrical calibration

Polymetron controller model under test

9500 / 9125

9125 (< V1.12) / 8925 / Others

8920

R∞

1

and 200 kΩ

2

NO

R∞

1

1

Cable disconnected from the sensor or sensor exposed to air

2

Use the precision resistance on the 9526

3. Conductivity calibration

Polymetron controller model under test

9500 / 9125 / 8920

9125 (< V1.12) / 8925

Others

K calculation (see

K Calculation

on page 18). On the controller

under test, enter the K Cell value

computed by the Polymetron 9526

Process: R∞ and comparative

measurement with 9526

1

Process: Comparative

measurement with 9526

1

1

Adjust the displayed value of the transmitter under test to that of the 9526 value

English 17