Specification

Details

Outputs

Analog output (temperature, conductivity/resistivity): 2 × 0/4-20 mA (linear, bilinear,
logarithmic) ± 0.1 mA

Alarms: 2 × thresholds or limits according to USP

Certifications

EN 61326-1: 2006; EN 61010-1: 2010

Sensor

Specification

Details

Sensor body material

Black PSU

Conductivity electrodes, internal and external

Stainless steel 316L

Cell constant K

0.01 (cm

-1

)

Conductivity range

0.01—200 μS.cm

-1

; Resistivity range: 5k Ω.cm—100 MΩ.cm

Maximum pressure

10 bar

Maximum temperature

125 °C (257 °F)

Accuracy

< 2%

Temperature response

< 30 seconds

Insulator

PSU

Connector

Glass polyester (IP65)

Controller

Specification

Details

Component description

Microprocessor-controlled and menu-driven controller that operates the sensor and
displays measured values.

Operating temperature

-20 to 60 ºC (-4 to 140 ºF); 95% relative humidity, non-condensing with sensor load
<7 W; -20 to 50 ºC (-4 to 104 ºF) with sensor load <28 W

Storage temperature

-20 to 70 ºC (-4 to 158 ºF); 95% relative humidity, non-condensing

Enclosure

1

NEMA 4X/IP66 metal enclosure with a corrosion-resistant finish

Power requirements

AC powered controller: 100-240 VAC ±10%, 50/60 Hz; Power 50 VA with 7 W
sensor/network module load, 100 VA with 28 W sensor/network module load
(optional Modbus, RS232/RS485, Profibus DPV1 or HART network connection).

24 VDC powered controller: 24 VDC—15%, + 20%; Power 15 W with 7 W
sensor/network module load, 40 W with 28 W sensor/network module load (optional
Modbus, RS232/RS485, Profibus DPV1 or HART network connection).

Altitude requirements

Standard 2000 m (6562 ft) ASL (Above Sea Level)

Pollution
degree/Installation
category

Polution Degree 2; Installation Category II

Outputs

Two analog (0-20 mA or 4-20 mA) outputs. Each analog output can be assigned to
represent a measured parameter such as pH, temperature, flow or calculated
values. Optional module supplies three additional analog outputs (5 total).

Relays

Four SPDT, user-configured contacts, rated 250 VAC, 5 Amp resistive maximum for
the AC powered controller and 24 VDC, 5A resistive maximum for the DC powered
controller. Relays are designed for connection to AC Mains circuits (i.e., whenever
the controller is operated with 115 - 240 VAC power) or DC circuits (i.e., whenever
the controller is operated with 24 VDC power).

4 English