Technical specifications – Hach-Lange POLYMETRON 83xx User Manual User Manual

Page 13

Advertising
background image

2 E

LECTRODES

C

ONDUCTIVITY

P

ROBE

- I

NSTRUCTION

M

ANUAL



621=183=000 Issue B

09/2010

Page 9

2. Technical

specifications

Model

8310 and 8315

8311 and 8316

8312 and 8317

8394

Applications :

Pure or high purity
waters

Meanly
conductived
solutions

Worm water,
process

Food and
pharmaceutical
industries
(sterilizable)

K (cm

-1

) 0.01

0.1

1

0.01

Precision K

< 2 %

< 2 %

< 2 %

< 2 %

Measurement
range (9125)

0.01 à 200 µS.cm

-1

0.1 µS à 2 mS.cm

-1

1 µS à 20 mS.cm

-1

0.01 à 200 µS.cm

-1

Temperature
response
Pt100 (t 90 %)

< 30 s

< 45 s

< 3 mn

< 45 s

Materials in contact
with the liquid:

Electropolished
stainless steel +
Psu (8310)

Electropolished
stainless steel +
PES 30 %
glassfiber + viton
(8315)

Stainless steel +
Psu (8311)

Stainless steel +
PES 30 %
glassfiber + viton
(8316)

Graphite + Psu
(8312)

Stainless steel +
graphite + viton
(8317)

Electropolished
stainless steel +
PEEK + EPDM
« FDA », Ra < 0.4

Temperature max
(°C)

125 (8310)

150 (8315)

125 (8311)

150 (8316)

125 (8312)

150 (8317)

150

P max (bars)

10 (8310)

25 (8315)

10 (8311)

25 (8316)

10 (8312)

25 (8317)

25

Process
connection

¾’’NPT ¾’’NPT ¾’’NPT Tri clamp 1.5 et 2’’