2 E
LECTRODES
C
ONDUCTIVITY
P
ROBE
- I
NSTRUCTION
M
ANUAL
621=183=000 Issue B
09/2010
Page 21
REMARK
The progressive accumulation of bubbles on
the surface of the probe:
¾ reduces its active surface,
¾ increases the cell constant,
¾ leads to a too low measure of conductivity.
Figure 3-7 - Circulation chamber
A
Very good installation:
Perfect immersion of the electrode surfaces.
B
Correct installation:
Satisfactory immersion of the electrode surfaces.
C Poor
installation:
Incomplete immersion of the electrodes, the
conductivity will be too low.
REMARK
The direction indicates the direction of flow.