2 E

LECTRODES

C

ONDUCTIVITY

P

ROBE

- I

NSTRUCTION

M

ANUAL



621=183=000 Issue B

09/2010

Page 21

REMARK

The progressive accumulation of bubbles on
the surface of the probe:
¾ reduces its active surface,
¾ increases the cell constant,
¾ leads to a too low measure of conductivity.

Figure 3-7 - Circulation chamber

A

Very good installation:
Perfect immersion of the electrode surfaces.

B

Correct installation:
Satisfactory immersion of the electrode surfaces.

C Poor

installation:

Incomplete immersion of the electrodes, the
conductivity will be too low.

REMARK

The direction indicates the direction of flow.