• 8394 :
This one installs perfectly in all short reduced
triclamp Tee "3A" Tri-clover
® starting from
1.5” (A), and also in short standard bent Ts at
90° "3A" starting from 2” (B).
Figure 3-6 - 8394 electrode
A
Very good installation:
Perfect immersion of the electrode surfaces.
B
Correct installation:
Satisfactory immersion of the electrode surfaces.
C Poor
installation:
Incomplete immersion of the electrodes, the
conductivity will be too low.
REMARK
The direction indicates the direction of flow.
In bypass
The flow-through chambers, are designed not
to retain air bubbles. To encourage the
extraction of the bubbles, use a minimum
flowrate of 20 l/hr (ideally 60 l/hr).