Page 20

09/2010

621=183=000 Issue B

8394 :

This one installs perfectly in all short reduced
triclamp Tee "3A" Tri-clover

® starting from

1.5” (A), and also in short standard bent Ts at
90° "3A" starting from 2” (B).

Figure 3-6 - 8394 electrode

A

Very good installation:
Perfect immersion of the electrode surfaces.

B

Correct installation:
Satisfactory immersion of the electrode surfaces.

C Poor

installation:

Incomplete immersion of the electrodes, the
conductivity will be too low.

REMARK

The direction indicates the direction of flow.

In bypass

The flow-through chambers, are designed not
to retain air bubbles. To encourage the
extraction of the bubbles, use a minimum
flowrate of 20 l/hr (ideally 60 l/hr).