Probe calibration
Whenever possible, we advise proceeding in
the following order:
• Temperature calibration
¾ It's an indispensable operation during
the commission for taking into account
the resistivity of the cable and offset
probe Pt100.
¾ Immerse the probe in a solution for
about 10 mn.
¾ Record the temperature of the solution
with a thermometer (precision < ± 0.1°C).
¾ Programm the transmitter in process
calibration mode.
¾ Adjust the value of the temperature read
with that of the thermometer.
• Conductivity calibration
First method (recommended) :
¾ Programm the transmitter in electrical
calibration mode. Choose the resistance
the closest possible to your process
(see table on the following page).
¾ First point: Remove the probe from the
liquid or unscrew the connector (Infinite
resistance, taking into account the
capacity of the cable).
¾ Second point: Connect the resistance
(precision <
0,1
%) of the value
programmed at the IN/OUT terminals of
the conductivity module.