Probe installation, On piping – Hach-Lange POLYMETRON 83xx User Manual User Manual
Page 21

2 E
LECTRODES
C
ONDUCTIVITY
P
ROBE
- I
NSTRUCTION
M
ANUAL
621=183=000 Issue B
09/2010
Page 17
Conductivity solution :
Resistivity solution :
0.1 µS.cm
-1
10 M
Ω.cm
10 µS.cm
-1
0.1 M
Ω.cm
1 mS.cm
-1
1 k
Ω.cm
10 mS.cm
-1
100
Ω.cm
R connected for K= 0.01 cm
-1
100 k
Ω 1
k
Ω
- -
R connected for K= 0.1 cm
-1
-
10 k
Ω 100
Ω
-
R connected for K= 1 cm
-1
-
100 k
Ω 1
k
Ω 100
Ω
- : unadapted measurement
Second method :
¾ Programm the transmitter in "process"
calibration mode.
¾ Make sure that the value recorded is
stabilised before adjusting it with that of
a precision calibration solution
(conductivity close to that of the
process).
Probe installation
On piping:
Entirely immerse the internal electrode in the
process (take into account the dimensions on
the following page in the event of a 90°
installation):