Conductivity solution :

Resistivity solution :

0.1 µS.cm

-1

10 M

Ω.cm

10 µS.cm

-1

0.1 M

Ω.cm

1 mS.cm

-1

1 k

Ω.cm

10 mS.cm

-1

100

Ω.cm

R connected for K= 0.01 cm

-1

100 k

Ω 1

k

Ω

- -

R connected for K= 0.1 cm

-1

-

10 k

Ω 100

Ω

-

R connected for K= 1 cm

-1

-

100 k

Ω 1

k

Ω 100

Ω


- : unadapted measurement

Second method :

¾ Programm the transmitter in "process"

calibration mode.

¾ Make sure that the value recorded is

stabilised before adjusting it with that of
a precision calibration solution
(conductivity close to that of the
process).