Electrodes: Check the insulation
between the two electrodes (R infinite
when probe exposed to air and dry).

Pay attention to the maintenance of
the probes!

E

The difficult conditions in which the
conductivity probes are often used makes
a periodic cleaning programm almost
obligatory.
This will contribute towards avoiding
accumulation at the surface of the
electrode of insulating layers leading to a
too weak reading of conductivity.
Ö In most uses, washing in hot water

with a household washing up liquid is
sufficient.

Ö

Greasy or oily layers can be
eliminated with methanol or ethanol.

Ö With solutions containing bacteria or

algae, use a chlorinated cleaning
product such as bleach.

Ö In the case of deposits of metallic

hydroxide, soak the probe for 10 mn in
a 20 % nitric acid solution.


F Carefully recalibrate the measurement

loop (See Chapter 3 - Calibration)