Analog output connections

W A R N I N G

Potential Electrocution Hazard. Always disconnect power to the instrument when making electrical
connections.

W A R N I N G

Potential electrocution hazard. In order to maintain the NEMA/IP environmental ratings of the
enclosure, use only conduit fittings and cable glands rated for at least NEMA 4X/IP66 to route cables in
to the instrument.

Two isolated analog outputs (1 and 2) are provided (

Figure 5

). Such outputs are commonly used for

analog signaling or to control other external devices.
Make wiring connections to the controller as shown in

Figure 5

and

Table 3

.

Note:

Figure 5

shows the back of the controller cover and not the inside of the main controller compartment.

Table 3 Output connections

Recorder wires

Circuit board position

Output 2–

4

Output 2+

3

Output 1–

2

Output 1+

1

1. Open the controller cover.
2. Feed the wires through the strain relief.
3. Adjust the wire as necessary and tighten the strain relief.
4. Make connections with twisted-pair shielded wire and connect the shield at the controlled

component end or at the control loop end.

• Do not connect the shield at both ends of the cable.
• Use of non-shielded cable may result in radio frequency emission or susceptibility levels higher

than allowed.

• Maximum loop resistance is 500 ohm.

5. Close the controller cover and tighten the cover screws.
6. Configure outputs in the controller.