Cleaning the controller

D A N G E R

Always remove power from the controller before performing maintenance activities.

Note: Never use flammable or corrosive solvents to clean any part of the controller. Use of these solvents may
degrade the environmental protection of the unit and may void the warranty.

1. Make sure the controller cover is securely closed.
2. Wipe the controller exterior with a cloth dampened with water, or with a mixture of water and mild

detergent.

Replace the resin

C A U T I O N

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local, regional and
national regulations.

The cartridge of resin can be replaced with a new cartridge, or as an alternative, keep the cartridge
and replace only the resin. In order to take advantage of the full system specifications, we
recommend the use of nuclear grade cationic resin.

1. Stop the sample flow using the sample flow adjustment valve.
2. Replace the cartridge or resin:

Cartridge—refer to

Resin cartridge installation

on page 9.

Resin—refer to

Figure 7

. Make sure to add deionized water at regular intervals when new resin

is added to compact the resin.

3. Open the degassing valve.
4. Open the sample flow adjustment valve and ensure everything is watertight and there are no

leaks.

5. Close the degassing valve when the measuring cell is free of air.
6. Set the sample flow to the required rate (between 5 and 20 L/h).
7. Reset the resin options. Refer to

Resin option

on page 19.