Specifications

Specifications are subject to change without notice.

Analyzer

Specification

Details

Dimensions

748 x 250 x 236 mm (29.4 x 9.8 x 9.3 in.)

Weight

7 kg (15.4 lb)

Sample flow rate

5—20 liters/hour

Ambient temperature 0—60 °C (32—140 °F)

Relative humidity

10—90%

Temperature sensor

Pt100

Accuracy

± 1% of displayed value; temperature < ± 0.2 °C

Calculated pH

Accuracy on conductivity measurement: ± 2%; Maximum difference calculated-theoretical
value: 0.1 pH

Display range

NH

3

; 7 < pH < 10; 2.8 μS/cm < C1 < 28 μS/cm; C2 < 0.5 μS/cm

NaOH; 7 < pH < 10.7; 2.5 μS/cm < C1 < 125 μS/cm; C2 < 100 μS/cm

Display resolution

Conductivity/resistivity: automatic point drift (minimum resolution 0.001 μS/cm) < 0.1 °C

Sample tubing

Polyethylene or PTFE or FEP; 0.2 to 6 bars (3 to 90 psi); 5 to 50 °C (40 to 120 °F); Input:
6-mm (standard) or 1/4-inch (with adapter); Output: 12-mm or 1/2-inch

Certifications

EN 61326-1: 2006; EN 61010-1: 2010

Sensor

Specification

Details

Sensor body material

Black PSU

Conductivity electrodes, internal and external

Stainless steel 316L

Cell constant K

0.01 (cm

-1

)

Conductivity range

0.01—200 μS.cm

-1

; Resistivity range: 5k Ω.cm—100 MΩ.cm

Maximum pressure

10 bar

Maximum temperature

125 °C (257 °F)

Accuracy

< 2%