4 menu structure, 1 sensor status, 2 sensor setup – Hach-Lange FP 360 sc User Manual User Manual
Page 18: 1 sensor status 4.4.2 sensor setup, Operation

18
Operation
4.4 Menu structure
Note: Refer to
for a list of all possible error and warning messages
together with a description of all necessary countermeasures to be taken.
4.4.1 SENSOR STATUS
SELECT SENSOR (if there is more than one sensor)
ERROR LIST
Possible error messages: SENSOR ERROR
WARNING LIST
Possible warning messages: TEST/MAINT, BULB CHANGE, LAST CONFIGUR, TARGET
VALUE
4.4.2 SENSOR SETUP
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
SET OUTMODE
Behavior of the outputs during calibration and zero point adjustment
HOLD
ACTIVE
SET TRANSFER
SENSOR MEASURE
Current, uncorrected measurement value
CONFIGURE
FACTOR: 0.1 to 100
OFFSET: –1000 to +1000
FACTOR
OFFSET
2 POINT
3 POINT
4 POINT
5 POINT
FACTOR
Is shown when FACTOR has been selected under CONFIGURE.
Refer to section
for detailed information.
OFFSET
Is shown when OFFSET has been selected under CONFIGURE.
Refer to section
for detailed information.
2 POINT
Is shown when 2 POINT has been selected under CONFIGURE.
Refer to section
for detailed information.
3 POINT
Is shown when 3 POINT has been selected under CONFIGURE.
Refer to section
for detailed information.
4 POINT
Is shown when 4 POINT has been selected under CONFIGURE.
Refer to section
for detailed information.
5 POINT
Is shown when 5 POINT has been selected under CONFIGURE.
Refer to section
for detailed information.
SET CAL DEFLT
Security query, reset to (FACTOR=1, OFFSET=0)