22

Operation

4.5.4.2 Adjust the zero point (OFFSET)

1. Open the MAIN MENU.

2. Select SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Select CALIBRATE and confirm.

5. Select CONFIGURE and confirm.

6. Press OFFSET and confirm.

7. Enter the required offset and confirm.

8. Go back to the MAIN MENU or the Measurement mode display.

4.5.4.3 Multi-point calibration (2 to 5-point calibration)

Note: Multi-point calibration means that the pairs must be input in ascending order.

1. Open the MAIN MENU.

2. Select SENSOR SETUP and confirm.

3. Select the corresponding sensor and confirm.

4. Select CALIBRATE and confirm.

5. Select CONFIGURE and confirm.

6. Select the type of calibration, e.g. 2 POINT and confirm.

7. Select 1PAIR and confirm.

8. Edit the TARGET VALUE and confirm.

9. Edit the ACTUAL VALUE and confirm.

10. Repeat the process for 2PAIR and confirm.

11. Go back to the MAIN MENU or the Measurement mode display.