Appendix A Modbus register

Table 1 Sensor Modbus registers

Register#

Data type

Length

R/W

Description

40001

Unsigned integer

1

R

Reserved

40002

Float

2

R

Measurement value as PAH, unit PPB

40004

Float

2

R

Measurement value as PAH, unit PPM

40006

Float

2

R

Measurement value as OIL, unit PPB

40008

Float

2

R

Measurement value as OIL, unit PPM

40016

String

6

R

Serial number

40022

String

8

R/W

Name of location

40030

Unsigned integer

1

R/W

Choose the parameter, 47=PAH, 48=OIL

40031

Unsigned integer

1

R/W

Measurement units

38=ppb, 39= g/L, 2=ppm, 0=mg/L

40040

Float

2

R/W

Offset for calibration

40042

Float

2

R/W

Factor for calibration

40050

Unsigned integer

1

R/W

Logging interval from 5 sec to 1800 sec

40051

Unsigned integer

1

R/W

Measuring interval from 1 sec to 300 sec

40052

Unsigned integer

1

R/W

Amplification for the low/high range probe,
0=AUTO,

1= 0.01 to 50/500 PPB,

2= 0.01 to 500/5000 PPB

40055

Unsigned integer

1

R/W

Set output mode for calibration

40056

Unsigned integer

1

R/W

Set output mode for service

40057

Float

2

R

Version of application file

40059

Float

2

R

The entry is for the application file

40061

Float

2

R

Version of the probe

40063

Unsigned integer

1

R

The entry is for the device driver file

40064

Unsigned integer

1

R

The entry is for the device driver file

40065

Unsigned integer

1

R

The entry is for the device driver file

40066

Integer

1

R

Days left until exchanging lamp,

negative values show that exchange is overdue

40067

Integer

1

R

Days left until maintenance,

negative values show that service is overdue

40068

Unsigned integer

2

R

Operating hours of analyzer