Specifications

Specifications are subject to change without notice.

Measurement

Measurement method

UV fluorescent measurement process for polycyclic aromatic
Hydrocarbons (PAH)

Excitation wavelength: 254 nm

Emission (measurement) wavelength: 360 nm

Measuring range

Low measurement range

0 to 50 ppb and 0 to 500 ppb in relation to PAH calibration standard,

corresponding to 0.1 to 1.5 ppm and 0.1 to 15 ppm of oil calibration
standard

High measurement range

0 to 500 ppb and 0 to 5000 ppb in relation to PAH calibration standard,

corresponding to 0.1 to 15 ppm and 0.1 to 150 ppm of oil calibration
standard

Display units

ppb, ppm, μg/L, mg/L

Reproducibility

2.5% of measurement value at a constant temperature

Measurement accuracy

5% of measurement value ±2% from measurement range limit

at a constant temperature

Limit of detection

1.2 ppb (PAH)

Response time

10 s (T90)

Calibration

Factory calibration with UV fluorescence calibration standard, custom
adaptation possible

Sensor software

Software version

From 1.14

Equipment properties

Weight

Measurement sensor

Stainless steel version 2.8 kg, titanium version 1.8 kg

Flow cell

Approximately 0.6 kg, including installation board approximately 2.0 kg

Pressure range

Sensor

Max. 30 bar

Flow cell

Max. 1 bar

Dimensions

Measurement sensor

(Ш Ч length)

68 mm x 306 mm (2.68 in. x 12.05 in.)

(without plugs or suspension pin)

68 mm x 399 mm (2.68 in. x 15.71 in.) (including suspension pin)

68 mm x 413 mm (2.68 in. x 16.26 in.)

(with additional cleaning option)

Flow cell (L × W × D)

98 mm x 98 mm x 150 mm (3.86 in. x 3.86 in. x 5.91 in.) (without
fittings),

installation board: 600 x 300 x 10 mm (23.62 in. x 11.81 in. x 0.39 in.)

Materials

Measurement sensor

Housing: stainless steel 1.4571 or titanium
Optic bracket: POM
Housing bolts: stainless steel 1.4571
Measurement window: synthetic quartz glass (Suprasil)
Gaskets (housing): Viton
Gaskets (measurement window): NBR (Nitrile Butadiene Rubber)

Flow cell

Housing: POM
Installation board: PVC
Gaskets: NBR (Nitrile Butadiene Rubber)
Fittings: nickel-plated brass

Shackle

Stainless steel 1.4301