36

Operation

5.3 [+DEVICEDATA] menu

Instrument parameters for adaptation to the related conditions.

Note: While the instrument is in this menu, sample pumping continues. If a key is not
pressed for longer than ten minutes, the instrument automatically returns to the display of
the data on operation.

*If the power supply is interrupted (power failure) both relay contacts are closed.

MENU

Command

Setting

Description

+DEVICEDATA

CONTRAST

-90 to +90

Reading angle for the display

LANGUAGE

D, GB, NL, F, I, E, PL, S

HEATER

ON, OFF, TIMER (time control)
Default setting: TIMER

Operation of the tube heaters

HEAT.START (only for [HEATER]:TIMER)

JAN to DEC,
Default setting: OCT

Activation of the heater at the start of the month selected

HEAT.STOP (only for [HEATER]:TIMER)

JAN to DEC,
Default setting: MRZ (March)

Deactivation of the heater at the end of the month selected

WARNING<

200...800 ml/h in 10 ml steps

Warning relay contact* set when the sample volume drops below
the amount set

FAULT<

100...600 ml/h in 10 ml steps

Fault relay contact* set when the sample volume drops below the
amount set

DATE

Setting for the date

TIME

Setting for the time

PASSWORD

4-digits with numbers 1-4

Password protection on activation

SW-VERSION

Only for information

Version of the instrument software

SW-CO.PROC.

Only for information

Version of the coprocessor software