Section 7

Faults, causes, rectification

Note: FILTRAX has three different types of error messages

7.1 Messages

Only text is displayed referring to the need for the
replacement of consumables - the green signal lamp
remains active.

7.2 Warnings

Text is displayed, both signal lamps are active and the
warning relay contact is set. If possible rectify without delay!

FAULT

CAUSE

RECTIFICATION

AIRFILTER COMP.

The operating time counter for the compressor air
filter [AIR F.C.] has expired

Replace compressor air filter and reset operating
time counter

AIRFILTER CASE

The operating time counter for the control unit air
filter [AIR F.H.] has expired

Replace control unit air filter and reset operating time
counter

PUMPWHEELS

The operating time counter for the pump rollers
[P.ROLL] has expired

Replace pump rollers and reset operating time
counter

PUMPCASSETTE

The operating time counter for the pump
cartridges [P.CARTR.] has expired

Replace pump cartridges and reset operating time
counter

COMPRESSOR

The operating time counter for the compressor
[COMP.] has expired

Replace compressor and reset operating time
counter

FAULT

CAUSE

RECTIFICATION

PUMPTUBES

The operating time counter for the pump tubes
[P.TUBES] has expired

Replace pump tubes and reset operating time
counter

LESS SAMPLE

The amount of sample from a module is below the
threshold set under [WARNING < ]

Check all tubes, if necessary clean filter modules

TEMPERATURE

The temperature in the control unit is too high, the
filter module venting is shut down automatically

Check fans in the control unit, if necessary replace
filter mats, reduce ambient temperature

TEST SETTINGS

On the [+DEVICEDATA] menu some settings may
have been lost

Check all instrument settings on the [+DEVICEDATA]
menu and [+SERVICE] menu