Section 6 maintenance, 1 maintenance work, Filtrax maintenance schedule – Hach-Lange FILTRAX User Manual User Manual

Page 44

Advertising
background image

42

Section 6

Maintenance

The manufacturer recommends taking out a service contract. This contract
extends the warranty to 5 years and ensures that all inspection and maintenance
work is performed by qualified personnel.

The maintenance to be performed by the user is then limited to regular visual
inspections, replacing consumables and cleaning, if necessary.

6.1 Maintenance work

The reliable and correct operation of this sample conditioning system can only be
ensured if the maintenance work is performed regularly as per the maintenance
calendar.

The necessary consumables are supplied or included in the set of annual
consumables.

Every 24 months
Replace
w Compressor

Weekly
Visual Inspection
w Sample Quality
w Device Function
w Air Filter
(Control Unit)

Every 3 months
Replace
w Pump Tubes
w Air Filter
(Control Unit)

Clean
w Filter Module
(if/as necessary)
w Sample Tubes
(if/as necessary)

Every 6 months
Check
w Air Filter
(Compressor)
w Pump Cartridges
w Pump Rollers
w Connecting Tubes
w Filter Module
w Elektronics

Every 12 months

Every 12 months

Replace

Replace

w Air Filter

w Air Filter

(Compressor)

(Compressor)
w Pump Cartridges

w Pump Cartridges
w Pump Rollers

w Pump Rollers
w Connecting Tubes

w Connecting Tubes

Number:

Place of sampling:

Installation Date:

by:

Inspection and Maintenance as part of the

Service Contract

Maintenance by the

User

DOC273.52.04006.Jul03

FILTRAX

Maintenance Schedule

N O T I C E

The sample pressure tube must never be clamped such that it is blocked off - the
pressure build up would either result in damage to the tubes or the tube connections
would spring off!!

Advertising