4. Connect the ground wire to the ground symbol terminal.
5. Connect the neutral wire to the NEUT terminal.
6. Connect the phase wire to the PHASE terminal.
7. Install the clear cover.
8. Close the enclosure cover. To maintain the enclosure environmental ratings, tighten the cover

screws. Make sure that there is a good seal.

9. Apply power to the instrument and the controller.

Table 1 Wiring for power

Terminal description

Wire color code—North America

Wire color code—Europe

Line (PHASE)

Black

Brown

Neutral (NEUT)

White

Blue

Protective Earth (ground symbol)

Green

Green with yellow tracer

Figure 6 Wiring schematic for 115 V

English 11