Installation, Instrument mounting – Hach-Lange High Output Airblast System (HOAB) User Manual

Page 6

Advertising
background image

Figure 1 Product components

1 Enclosure

8 Lock washer

2 Voltage barrier (sc100)

9 Bolt, hex head, slotted

3 Screw, barrier hold-down

10 Nut, hex

4 Voltage barrier (sc200)

11 Bracket

5 Tab, wall mounting

12 Bracket

6 Screw, flat head, slotted

13 Uni-strut

7 Screw, pan head, slotted

Installation

D A N G E R

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.

Instrument mounting

N O T I C E

If the instrument is installed in direct sunlight, the operating temperature might rise above its specified limit and
the compressor will not start. Install the instrument in an indoor or outdoor location that is not in direct sunlight.

1. Install the instrument near the installed sensor (

Figure 5

). Do not extend the air delivery tubing

beyond 7.6 m (25 ft). This keeps the capacity of the system the same and does not decrease the
response time for the air delivery.

2. Mount the instrument in a location where the ambient temperature will not be more than the

instrument temperature limits. Refer to

Specifications

on page 3.

6 English

Advertising