Troubleshooting, Replacement parts, Figure 9 – Hach-Lange High Output Airblast System (HOAB) User Manual

Page 16

Advertising
background image

Figure 9 Fuse replacement

1 Protective cover

3 Input fuse (2x)

2 Power switch

4 Controller relay fuse

Troubleshooting

Problem

Possible cause

Solution

The compressor operates but
no air comes from the
instrument.

The air tubing is pinched
or damaged.

Examine the air tubing for sharp bends and/or
pinched areas. Repair or replace the air tubing as
necessary.

The air tubing is
obstructed.

Examine the air tubing for obstructions. Clean or
replace the air tubing as necessary.

The compressor does not
operate.

A fuse(s) is open.

Examine the wiring to see if there is a wiring problem.
If not, replace any bad fuse(s). Refer to

Replace the

fuses

on page 15.

The compressor does not
operate on schedule.

There is a controller
problem.

Make sure that the controller has the correct
programming. Make sure that the controller is
operating correctly.

Replacement parts

Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.

Description

Item no.

Fuse, 5 A, 250 V slow-blow (230 V model)

4693800

Fuse, 8 A, 250 V slow-blow (115 V model)

6172000

Fuse, 0.25 A, 250 V slow-blow (115/230 V models)

5966073

Mounting kit (includes mounting channels, screws, washers, label, clamps)

6864900