Connect a serial device (optional)

One RS232 connection is available on the J4 connector. Connect a
printer or the serial port of a computer (COM 1 or COM 2) to the
RS232 connection to send the buffered data to a printer or a computer.
Refer to

Send the data log to a printer or computer

on page 15.

If the optional RJ-45 to DB-9 converter is used, cut an RJ-45 cable and
then separate and identify the wires. Refer to

Figure 5

. Refer to

Table 4

for the RS232 wiring connections.

Note: The RS232 connection is not available when the digital input (CTS pin) is
used.

Figure 5 RJ-45 to DB-9 converter wiring

1 To PC or printer

2 RJ-45 to

DB-9 converter
(SA000070-01)

3 RJ-45 cable (user

supplied)

Table 4 RS232 connections

DB-9 pin

RJ-45 pin

J4 pin

Pin 2 RX

Pin 3

TX

Pin 3 TX

Pin 4

RX

Table 4 RS232 connections (continued)

DB-9 pin

RJ-45 pin

J4 pin

Pin 4 DTR

Pin 5

CTS

Pin 5 GND

Pin 8

GND

Supply power to an external device (optional)

+5 VDC or +12 VDC is available on JP1 to supply power to an external
device. +5 VDC is supplied when the jumper is across the +5V pin and
pin 2 of JP1. +12 VDC is supplied when the jumper is across the Vps pin
and pin 2 of JP1.

Note: The V+ pin and GND pin on the J1 connector are for Service use only.

Connect units in a network (optional)

Multiple units can be connected together in a network. Use the
RS485 connection of each unit and 19 AWG, twisted shielded pair wire
to connect the units together.
Notes:

• A maximum of 32 units in a network can be connected to one

RS485 network.

• The host computer must have data acquisition software.
• The communication protocol is Modbus. Refer to the WPC Modbus

documentation on the manufacturer's website.

1. On the first unit:

a. Push the incoming cable from the RS232/RS485 converter

through a strain-relief fitting.

b. Push the outgoing cable through a strain relief fitting.
c. Connect the incoming and outgoing cable wires to the

J4 connector. Refer to

Table 5

.

2. On the second unit:

a. Push the incoming cable from the first unit through a strain-relief

fitting.

b. Push the outgoing cable through a strain relief fitting.