Specifications

Specifications are subject to change without notice.

Specification

Details

Dimensions (W x D x H)

248 x 114 x 302 mm (9.75 x 4.50 x 11.88 in.) with
sensor and mounting brackets

Enclosure

NEMA 4X

Light source

780 nm, 5 mW class 3B laser diode; 30,000 hours
lifetime

Detector

Photodiode

Display

LCD, 4-line x 16-character

Port sizes

Two compression fittings for 0.25-in. OD inlet and
outlet tubing

Wetted materials

Fused silica, Viton (fluorocarbon) and Kynar (PVDF)

Flow cell

WPC-21: 600 x 600 µm fused silica with AR
coatings
WPC-22: 800 x 800 µm fused silica with AR
coatings

Weight

2.25 kg (5 lb)

Installation category

II

Protection class

I

Pollution degree

2

Operating environment

WPC-21: 5 to 40 °C (41 to 104 °F), 5 to 90% relative
humidity, non-condensing
WPC-22: 5 to 45 °C (41 to 113 °F), 5 to 90% relative
humidity, non-condensing

Storage environment

–20 to 60 °C (–4 to 140 °F), 98% relative humidity,
non-condensing (maximum)

Altitude

2000 m (6562 ft) maximum

Power requirements

230 VAC ±15% , 20 VA, 50/60 Hz

Specification

Details

Detection method

Light blocking

Calibration

WPC-21: Factory calibrated at flow rate of
50 mL/minute, NIST traceable, eight particle sizes
WPC-22: Factory calibrated at flow rate of
100 mL/minute, NIST traceable, eight particle sizes

Particle sizes

WPC-21: 1.3, 2, 3, 5, 7, 10, 15 and 25 µm
WPC-22: 2, 5, 7, 10, 15, 25, 50 and 100 µm

Coincidence fields

WPC-21: 10% loss at 25,000 particles/mL
WPC-22: 10% loss at 15,000 particles/mL

Number of channels

Two channels; Channel 1: smallest particle size,
Channel 2: one of the other seven channel sizes
(fixed) or all seven particle sizes (scrolling)

Measurement units

Number of particles per mL

Sensor resolution

Less than 10% at 10 µm as described in ASTM-
F658-87

Sensor counting efficiency Meets JIS 9925 B-1997 standards

Sample pressure

8.3 bar (120 psi) maximum

Sample flow rate

WPC-21: 45–55 mL/minute
WPC-22: 90–110 mL/minute

Sample temperature

0 to 50 °C (32 to 122 °F), non-freezing

Digital input

One; 0–10 VDC, ± 24 VDC < 10 mA

Digital outputs

Two; open collector (100 mA, 24 VDC maximum)

Analog inputs

Two; 0–10 VDC

Analog outputs

2 channels: 4–20 mA

Communications

RS485 (protocol: MODBUS) or RS232, 19 AWG
twisted shielded pair wire

English 3