Connect the analog outputs (optional)

Two 4–20 mA analog outputs (AO1 and AO2) are available at the
J3 connector. The analog output signal is proportional to the counts/mL
of the channel. The counts/mL for Channel 1 is represented on AO1.
The counts/mL for Channel 2 is represented on AO2.
Connect a current receiving device to each analog output with
24–28 AWG wire. Refer to

Table 1

.

If the current receiving device(s) has a load resistor of 500 ohms or
more, supply external voltage (40 VDC maximum) to Vext– and Vext+ of
the J1 connector to supply sufficient voltage to the current receiving
device.

Table 1 Analog output connections (4–20 mA)

Pin

Description

AO2–

Neutral

AO2+

Positive

AO1–

Neutral

AO1+

Positive

Connect the analog input (optional)

Two analog inputs (AI1 and AI2) are available on the J3 connector.
Connect auxiliary instruments (i.e., flow meters, turbidity meters,
pressure transducers, pH meter) to the analog inputs to show the
voltage supplied by the auxiliary instruments on the instrument display
(0–10 VDC). Refer to

Identify the analog input values

on page 15.

Connect an auxiliary instrument to each analog input with 22–24 AWG
wire. Use 19 AWG wire instead if the wire is 7.62 m (25 ft) long or more.
Refer to

Table 2

.

Table 2 Analog input connections (0–10 VDC)

Pin

Description

AI2–

Neutral

AI2+

Positive

Table 2 Analog input connections (0–10 VDC) (continued)

Pin

Description

AI1–

Neutral

AI1+

Positive

Connect the digital input (optional)

One digital input is available on the J4 connector. Connect a flow
monitoring device to the digital input (CTS pin) to trigger a Flow alarm
when the voltage on the digital input is high or low, depending on the
Flow Alarm State setting. Refer to

Configure the particle counter

on page 13 to configure the Flow Alarm State setting and enable the
Flow alarm.
Connect the flow monitoring device to the CTS pin with 24–28 AWG
wire.

Note: The digital input is not available when the RS232 connection is used to
connect a printer or a computer.

Connect the digital outputs (optional)

Two digital outputs (ALM and CAL) are available on the J1 connector.
Refer to

Table 3

. Connect a compatible device such as a remote alarm

indicator or buzzer to the applicable digital output with 24–48 AWG wire.
The ALM pin and CAL pin are open collector outputs that supply a
ground connection when energized. Obey the digital output ratings in

Specifications

on page 3.

Table 3 Digital output descriptions

Digital output Description

ALM pin

Energizes when a Flow alarm or Count alarm is active. Refer to

Configure the particle counter

on page 13 to configure the

Flow alarm and Count alarm.

CAL pin

Energizes when a high or low calibration error is active.

English 9