Plumbing, Plumb the instrument, Sample line guidelines – Hach-Lange WPC-22 User Manual

Page 11: J4 connector. refer to, Table 5

Advertising
background image

c. Connect the incoming and outgoing cable wires to the

J4 connector. Refer to

Table 5

.

3. Do step 2 again for the third and remaining units until all the units

have incoming and outgoing connections. The last unit will not have
an outgoing cable.

Note: In a network installation, the outgoing cable from the previous unit is
always the incoming cable on the next unit.

4. On the last unit, install a jumper across JP2. Do not install a jumper

across JP2 on the other units.

5. Set the MODE option to N (network mode). Refer to the Date/MODE

option in

Configure the particle counter

on page 13.

6. Assign a unique address to each unit. Refer to the Unit ID option in

Configure the particle counter

on page 13.

Table 5 RS485 connections

Wire

J4 pin

RS485A

+

RS485B

Shield

SGND

Plumbing

Plumb the instrument

Items to collect:

• 6 mm OD opaque tubing
• Two, 6 mm ferrules

Select tubing based on the worst case environmental conditions (e.g.,
temperature range, exposure to sunlight, contaminants in the air) and
the highest possible inlet pressure (maximum of 120 psi).

1. Remove the tubing from the sample line fitting and drain line fitting

and discard. Do not reuse the tubing for the sample line or drain line.

2. Connect a sample line to the sample line fitting on the body of the

instrument. Refer to

Figure 6

.

3. Connect a drain line to the drain line fitting on the body of the

instrument. Refer to

Figure 6

.

4. Install a Y-strainer in the sample line as necessary to prevent large

particles from entering the sensor which will result in a blockage.

5. Install a manual isolation valve in the sample line so that the flow to

the instrument can be stopped as necessary.

Figure 6 Plumb the instrument

1 Sample line fitting

3 Plastic nut

2 Drain line fitting

4 Ferrule, 6 mm

Sample line guidelines

Select a good, representative sampling point for the best instrument
performance. The sample must be representative of the entire system.
To prevent erratic readings:

• Collect samples from locations that are sufficiently distant from points

of chemical additions to the process stream.

• Make sure that the samples are sufficiently mixed.
• Make sure that all chemical reactions are complete.

English 11

Advertising
This manual is related to the following products: